Life Air Series Specification

Specifications Life Air 100 series Life Air 200 series
 
General
Intended Use Supply dry filtered air for respiratory ventilator or anesthesia machine
Area Intensive care, recovery room
Type of operation Long-term use, applicable as main or back-up supply
Noise level ≤ 50 dB(A) at distance of 1m
Over Pressure Relief 6.0 bar / 87 psi
Electrical Protection Mains power fuse with ON/OFF switch
Mounting Trolley with castor wheels
Service Mode -- Available
Weight < 33 kg < 35 kg
Dimensions W 42 cm, D 31 cm, H 32 cm W 49 cm, D 31 cm, H 32 cm
Power Supply
Power Supply 220 to 240 V AC, 50 Hz, single phase
Supply range 198 to 264 V AC
Power Consumption < 550 W at nominal voltage
Power inlet cable 230 V / 50 Hz, 15 A, single phase, 3 mtrs
Compressor Capacity
Outlet flow (continuous) 40 L/min (expanded to ambient air) at 50 psi / 3.5 bar
Outlet flow (peak) 200 L/min (with pressure drop of 7 psi / 0.5 bar)
Inlet pressure 2.5 to 6.0 bar / 36 to 87 psi
Outlet pressure 3.0 to 4.0 bar / 44 to 58 psi
Built-in air reservoir 1.25 L
Particulate filtration 5 micron
Dew point suppression > 5 °C / 9 °F below ambient temperature at 50 psi / 3.5 bar NA
Alarms
High Temperature Alarm ACOUSTIC
85 dB(A) max. at distance of 1 m
OPTICAL
Alarm message in LCD with blinking Red LED
NA
Low / high pressure alarm ACOUSTIC
85 dB(A) max. at distance of 1 m
OPTICAL
Alarm message in LCD with blinking Red LED
NA
Monitoring
Outlet pressure 0.0 to 9.9 bar / 0 to 99 psi Pressure gauge (0 to 10 bar)
Operating time 0 to 99999 Hours Hourmeter (0 to 99999 Hours)
Display 2 line matrix display NA
Ambient temperature 0 to 99 °C / 0 to 999 °F NA
Cooler temperature 0 to 99 °C / 0 to 999 °F NA
Pump temperature 0 to 99 °C / 0 to 999 °F NA
AC input 180 to 300 VAC NA
Cooler status ON or OFF NA
Water drain status ON or OFF NA
Power ON indicator Green LED NA
Alarm indicator Red LED NA
Environmental conditions – Operations
Ambient temperature +10 to +40°C (50 to 104°F)
Ambient relative humidity 15 to 95%
Atmospheric pressure 660 to 1200 hPa (495 to 900 mm Hg)
Environmental conditions – Transportation & storage
Ambient temperature -40 to +70°C (-40 to 158°F)
Ambient relative humidity 10 to 95%
Atmospheric pressure 500 to 1200 hPa (375 to 900 mm Hg)

*In view of continuous improvement, specifications are subject to change without prior notice.